Double Beam UV/VIS Spectrophotometer

      UV-4802 is a double beam design. It is a stand-alone model with 1.8nm fixed bandwidth and has all the features that SpectroQuest line offers in a stand-alone unit. The two detectors are measuring sample and reference respectively and simultaneously for optimizing measurement accuracy. It provides excellent performance for measurements in the range of 190nm to1100nm.It is suitable for pharmaceutical, biochemical and clinical lab applications as well as routine applications such as quantitative analyses, kinetics, spectrum scanning, multiple components and DNA/Protein. Optional PC download software and PC Windows® application software make this instrument versatile


SHARE:

Specifications

Wavelength Range :190-1100nm

Spectral Bandwidth :1.8nm

Optical System : DoubleBeam,Grating1200lines/mm

Wavelength Accuracy:±0.3nm

Wavelength Resolution: ±0.1nm

Wavelength Repeatability:≤0.1nm

Stability:±0.0005Abs@500nm

Baseline Flatness: ±0.001A

Display :Graphic LCD(320×240)dot

Power Requirements :220V-240V/50HZ,110V-120V/60HZ Switchable

Dimensions(W*D*H) :620*400*280mm

Weight :24Kg

Shipping Dimensions :720*525*370mm

Shipping Weight :28KG


PREV: NOTHING